Methods for Chemical Analysis of Water and Wastes

PDF Publication Title:

Methods for Chemical Analysis of Water and Wastes ( methods-chemical-analysis-water-and-wastes )

Previous Page View | Next Page View | Return to Search List

Text from PDF Page: 226

Instrument Parameters 1 DryIng tune and temperature 30 sec @ 125°C 2 CharnngtImeandtemperature 30sec@ 1200°C 3. AtomIzIng tIme and temperature 10 sec @ 2700°C 4 Purge Gas Atmosphere Argon 5 Wavelength 1960nm 6 Other operatIng parameters should be set as speCIfied by the partICular Instrument manufacturer AnalysIs Procedure 1 For the analysIs procedure and the calculatIon see "Furnace Procedure" part 9 3 of the AtomIC AbsorptIOn Methods sectIOn oftms manual Notes 1. The above concentratIon values and Instrument condItIOns are for a PerkIn-Elmer HGA- 2100, based on the use of a 20 ul lUJecnon, purge gas lUterrupt and non-pyrolync graphIte Smaller sIze furnace deVIces or those employIng faster rates of atomIzatIOn can be operated USIng lower atoffilzatlOn temperatures for shorter tIme penods than the above recommended settIngs 2. The use ofbackground correctIon IS recommended 3 Selemum analysIs suffers Interference from chlondes (> 800 mg/l) and sulfate (> 200 mg/l) FortheanalysIsofIndustnaleffiuentsandsampleswIthconcentratIonsofsulfate from 200 to 2000 mg/l, both samples and standards should be prepared to contaIn 1% mckel 4 For every sample matnx analyzed, venficatlOn IS necessary to deterrmne that method of standard addItIOn IS not reqUIred (see part 5 2 1 of the AtomIC AbsorptIOn Methods sectIon ofthIS manual) 5 For qualIty control reqUIrements and optIOnal recommendatIons for use In dnnkIng water analyses, see part 10 of the AtomIC AbsorptIOn Methods sectIOn of thIs manual 6 If method of standard addItIOn IS reqUIred, follow the procedure given earlIer In part 8 5 ofthe AtomIC AbsorptIOn Methods sectIOn ofthIS manual 7 Data to entered Into STORET must be reported as ug/l Precision and Accuracy 1 USIng a sewage treatment plant effiuent contaInIng <2 ug/l and spIked wIth a concentratIon of20 ug/l, a recovery of99% was obtaIned 2 USIng a senes of Industnal waste effiuents spIked at a 50 ug/l level, recovenes ranged from 94 to 112% 3 USIng a 0 1% mckel mtrate solutIOn as a synthetIc matnx with selemum concentratIOns of 5, 10,20,40, 50, and 100 ug/l, relative standard deVIatIOns of 14 2, 11 6, 9 3, 72, 64 and 4 1%, respectively, were obtaIned at the 95% confidence level 2702-2 " "

PDF Image | Methods for Chemical Analysis of Water and Wastes

methods-chemical-analysis-water-and-wastes-226

PDF Search Title:

Methods for Chemical Analysis of Water and Wastes

Original File Name Searched:

epa600_4_79_020.pdf

DIY PDF Search: Google It | Yahoo | Bing

CO2 Organic Rankine Cycle Experimenter Platform The supercritical CO2 phase change system is both a heat pump and organic rankine cycle which can be used for those purposes and as a supercritical extractor for advanced subcritical and supercritical extraction technology. Uses include producing nanoparticles, precious metal CO2 extraction, lithium battery recycling, and other applications... More Info

Heat Pumps CO2 ORC Heat Pump System Platform More Info

CONTACT TEL: 608-238-6001 Email: greg@infinityturbine.com | RSS | AMP